Expanded Measurement Uncertainty

Expanded measurement uncertainty is obtained by multiplying combined standard uncertainty by a coverage factor. It provides an uncertainty interval intended to cover a specified proportion of the distribution of values reasonably attributable to the measured quantity, subject to the stated statistical assumptions.

\(U = k u_{\mathrm{c}}(y)\)

What this formula means

Expanded measurement uncertainty is obtained by multiplying combined standard uncertainty by a coverage factor. It provides an uncertainty interval intended to cover a specified proportion of the distribution of values reasonably attributable to the measured quantity, subject to the stated statistical assumptions.

Scientific principle

Principle

Combined standard uncertainty describes the standard uncertainty of a measurement result. Expanded uncertainty increases this value using a coverage factor to express a wider interval associated with a chosen level of coverage.

Derivation

  • 1. Determine the combined standard uncertainty u_c(y) from the relevant uncertainty contributions.

  • 2. Select a coverage factor k appropriate to the intended coverage probability and statistical model.

  • 3. Multiply the combined standard uncertainty by the coverage factor.

  • 4. The expanded uncertainty is U = k u_c(y).

  • 5. The measurement result may then be reported as y ± U, with the coverage interpretation stated.

Final Result

U = k u_c(y)

Reporting Relation

Measurement result = y ± U

Common Approximation

For an approximately normal distribution with sufficiently large effective degrees of freedom, k ≈ 2 corresponds to approximately 95% coverage.

Variables and their meanings

SymbolMeaningUnit

U

Expanded measurement uncertainty

Same unit as y

k

Coverage factor

Dimensionless

u_c(y)

Combined standard uncertainty

Same unit as y

y

Measurement result

Unit of the measured quantity

SI units and conversions

Expanded Uncertainty

Same unit as the measurement result

Coverage Factor

Dimensionless

Combined Standard Uncertainty

Same unit as the measurement result

Conditions of use

  • Combined standard uncertainty must be evaluated first.

  • The coverage factor must be selected according to the intended coverage probability and distribution assumptions.

  • The reporting convention should state the coverage factor.

  • The statistical model should be appropriate for the measurement process.

  • If the effective degrees of freedom are limited, an appropriate Student's t coverage factor may be needed.

Limitations and assumptions

  • A coverage factor of 2 does not guarantee exactly 95% coverage in every situation.

  • Coverage depends on the probability distribution and effective degrees of freedom.

  • Expanded uncertainty is not the same as the maximum possible error.

  • The interval y ± U is not automatically a strict bound containing the true value.

  • Unrecognized systematic effects cannot be corrected simply by increasing the coverage factor.

Step-by-step worked examples

ProblemSolutionAnswer

A measurement has a combined standard uncertainty of 0.25 cm. Calculate its expanded uncertainty using a coverage factor of 2.

  • Given: u_c(y) = 0.25 cm.

  • Coverage factor: k = 2.

  • Use: U = k u_c(y).

  • U = 2 × 0.25 cm.

  • U = 0.50 cm.

The expanded uncertainty is 0.50 cm. The result can be reported as y ± 0.50 cm, with k = 2 stated.

A calibrated instrument gives a measurement result of 15.80 V with combined standard uncertainty 0.12 V. Calculate the expanded uncertainty for k = 2.

  • Given: y = 15.80 V.

  • u_c(y) = 0.12 V.

  • k = 2.

  • U = 2 × 0.12 V = 0.24 V.

  • Reported interval: 15.80 V ± 0.24 V.

Expanded uncertainty = 0.24 V. The reported result is 15.80 V ± 0.24 V, with coverage factor k = 2.

Real-world applications

  • Calibration certificates

  • Laboratory measurement reporting

  • Industrial metrology

  • Scientific instrument specifications

  • International measurement comparisons

Formula Meaning

Expanded measurement uncertainty is obtained by multiplying combined standard uncertainty by a coverage factor. It provides an uncertainty interval intended to cover a specified proportion of the distribution of values reasonably attributable to the measured quantity, subject to the stated statistical assumptions.

Sources and further reading

No source has been supplied for this item yet. Treat the claim as needing editorial verification.