The coverage factor is the ratio of expanded measurement uncertainty to combined standard uncertainty. It determines how many combined-standard-uncertainty units are used to construct an expanded uncertainty interval.
Coverage Factor Relation
The coverage factor is the ratio of expanded measurement uncertainty to combined standard uncertainty. It determines how many combined-standard-uncertainty units are used to construct an expanded uncertainty interval.
\(k = U/u_{\mathrm{c}}(y)\)
What this formula means
Scientific principle
Principle
A measurement result is often accompanied by an interval intended to provide a specified coverage probability. The coverage factor scales the combined standard uncertainty to obtain the expanded uncertainty. Its appropriate value depends on the probability distribution and effective degrees of freedom.
Derivation
1. Expanded uncertainty is defined as U = k u_c(y).
2. Divide both sides by the combined standard uncertainty u_c(y), assuming it is nonzero.
3. Rearranging gives k = U/u_c(y).
4. The coverage factor is dimensionless because the numerator and denominator have the same unit.
5. For an approximately normal distribution with sufficiently large degrees of freedom, k near 2 is commonly associated with approximately 95% coverage.
6. For finite effective degrees of freedom, the Student's t distribution can provide a more appropriate factor.
Final Result
k = U/u_c(y)
Expanded Uncertainty Relation
U = k u_c(y)
Normal Distribution Approximation
k ≈ 1.96 for an exact central 95% interval under a standard normal model
Common Practical Approximation
k ≈ 2 for approximately 95% coverage when the normal approximation is suitable
Variables and their meanings
| Symbol | Meaning | Unit |
|---|---|---|
k | Coverage factor | Dimensionless |
U | Expanded measurement uncertainty | Unit of measured quantity |
u_c(y) | Combined standard uncertainty | Same unit as U |
ν_eff | Effective degrees of freedom | Dimensionless |
SI units and conversions
Coverage Factor
Dimensionless
Expanded Uncertainty
Same unit as measurement result
Standard Uncertainty
Same unit as measurement result
Conditions of use
Expanded uncertainty and combined standard uncertainty must refer to the same measurement result.
Combined standard uncertainty must be nonzero.
The coverage factor must be interpreted together with the statistical distribution and coverage objective.
For finite effective degrees of freedom, an appropriate t-distribution quantile may be required.
The coverage probability should be stated when reporting the expanded uncertainty.
Limitations and assumptions
The coverage factor alone does not determine coverage probability without a distributional model.
The commonly used value k = 2 is an approximation, not a universal 95% guarantee.
A small effective degree of freedom can require a coverage factor greater than 2 for a central 95% interval.
The coverage factor is not an instrument accuracy rating.
The equation is undefined if combined standard uncertainty is zero.
Step-by-step worked examples
| Problem | Solution | Answer |
|---|---|---|
A measurement has expanded uncertainty 0.60 mm and combined standard uncertainty 0.30 mm. Calculate the coverage factor. |
| The coverage factor is 2. |
A measurement has combined standard uncertainty 0.08 s and uses a coverage factor of 1.96. Calculate its expanded uncertainty. |
| The expanded uncertainty is 0.1568 s, approximately 0.157 s. |
Real-world applications
Determining expanded uncertainty
Calibration and measurement reporting
Statistical confidence and coverage analysis
Metrology documentation
Laboratory quality assurance
Formula Meaning
The coverage factor is the ratio of expanded measurement uncertainty to combined standard uncertainty. It determines how many combined-standard-uncertainty units are used to construct an expanded uncertainty interval.