In Doppler ultrasonography, a probe sends ultrasound into the body and receives echoes from moving scatterers, especially red blood cells. Motion toward or away from the probe changes the frequency of the reflected signal. In the common pulse-echo model, the factor of two accounts for the Doppler shift occurring during both the outgoing and returning paths. The frequency shift depends on the transmitted frequency, target velocity, speed of sound in tissue and angle between the ultrasound beam and the direction of motion. The measured shift can help estimate the component of blood velocity along the beam. The method is widely used to assess blood flow, but the angle and measurement conditions must be considered when interpreting results.
Doppler Ultrasonography
Doppler ultrasonography uses frequency shifts in reflected ultrasound to estimate the motion and speed of blood or other moving targets.
Δf = 2f₀v_b cosθ/c
Explanation
Variables
| Symbol | Meaning | Unit |
|---|---|---|
Δf | Measured Doppler frequency shift | Hz |
f₀ | Transmitted ultrasound frequency | Hz |
v_b | Velocity of the moving blood cells along their direction of motion | m/s |
θ | Angle between ultrasound beam and blood-flow direction | Degrees |
c | Speed of sound in tissue | m/s |
Example or application
A Doppler ultrasound system transmits a 5 MHz signal. If blood moves toward the probe at 0.5 m/s, the beam-flow angle is 60°, and tissue sound speed is 1540 m/s, the ideal shift magnitude is approximately 2 × 5,000,000 × 0.5 × cos60° / 1540 ≈ 1623 Hz.
Key Points
Doppler ultrasound detects frequency shifts from moving scatterers.
Red blood cells are common sources of reflected ultrasound.
The pulse-echo model includes a factor of two.
The beam-flow angle affects the measured frequency shift.
Angle and instrument settings influence velocity estimates.
Applications
Assessment of blood-flow direction and speed
Evaluation of blood vessels
Cardiac Doppler examinations
Detection of abnormal flow patterns
Monitoring circulation
Common Mistakes
Ignoring the angle between the beam and blood flow.
Forgetting the factor of two in the usual reflected pulse-echo model.
Treating the measured velocity component as the full blood velocity when the angle is not zero.
Learning Outcomes
Explain the basic principle of Doppler ultrasonography.
Identify the source of reflected ultrasound signals.
Explain the role of the beam-flow angle.
Interpret the factors that determine the Doppler frequency shift.
Practice Questions
| Question | Answer |
|---|---|
What moving structures commonly produce Doppler echoes in blood-flow ultrasound? | Moving red blood cells. |
Why does the usual pulse-echo formula contain a factor of two? | The sound experiences a Doppler shift on the outgoing path and again on the reflected return path. |
What happens to the measured Doppler shift when the beam is perpendicular to blood flow in the ideal model? | The shift is zero because cos 90° = 0. |
Related Concepts
Doppler Effect in Sound Waves
Doppler Shift in Sound Waves
Acoustic Doppler Measurement
Medical and Diagnostic Acoustics
Sources and verification
Sources are shown as provided in the project data. Not every source has yet been independently verified. Please report errors or missing citations.
- Diagnostic Ultrasound: Physics and Equipment — Peter R. Hoskins, Kevin Martin and Abigail Thrush
- Fundamentals of Acoustics — Lawrence E. Kinsler et al.