An ion-selective electrode (ISE) responds preferentially to a particular ion through a selective membrane. The membrane may be glass, a solid crystalline material, or a liquid membrane containing an ion-selective carrier. A potential develops across the membrane because of differences in ion activity between the sample and the internal reference system. The potential is measured against a reference electrode and related to the logarithm of the target ion activity. Selectivity is not absolute: other ions can contribute to the measured response. The Nicolsky–Eisenman equation extends the ideal Nernst response by accounting for interfering ions.
Ion-Selective Electrodes
Ion-selective electrodes are electrochemical sensors that develop a potential related to the activity of a particular ion in a sample.
Explanation
Variables
| Symbol | Name | Description |
|---|---|---|
E | Electrode potential | Measured potential associated with the ion-selective electrode system. |
E° | Constant potential term | Includes standard and system-dependent potential contributions. |
R | Gas constant | 8.314 J mol⁻¹ K⁻¹. |
T | Absolute temperature | Temperature in kelvin. |
z | Ion charge number | Signed charge number of the relevant ion. |
F | Faraday constant | Approximately 96485 C mol⁻¹. |
a | Ion activity | Dimensionless relative activity of the target ion. |
K(A,B) | Selectivity coefficient | Describes the relative response to interfering ion B. |
Example or application
A worked example has not yet been added.
Definition
An ion-selective electrode is an electrochemical sensor containing a membrane that produces a potential dependent primarily on the activity of a selected ion.
Purpose
To measure the activity or, after suitable calibration and sample treatment, estimate the concentration of a specific ion in a solution.
Importance
Ion-selective electrodes enable rapid measurements of ions in environmental, clinical, agricultural, and industrial samples. They are especially useful when direct measurement of an ionic species is preferable to a more complex analytical procedure.
Key Concepts
Selective membrane: Membrane that preferentially interacts with a target ion.
Membrane potential: Potential difference developed across the selective membrane.
Ion activity: Effective concentration of an ion accounting for non-ideal solution behavior.
Nernst response: Logarithmic relationship between potential and ion activity.
Selectivity coefficient: Parameter describing the relative influence of an interfering ion.
Internal reference system: Provides a stable internal potential.
Detection limit: Lowest analyte activity that can be reliably distinguished under specified conditions.
Principle
The membrane selectively interacts with the target ion and establishes a potential related to its activity. At a fixed temperature and under controlled conditions, the potential commonly changes approximately linearly with the logarithm of ion activity.
Working Process
Choose an ion-selective electrode suitable for the target ion.
Prepare calibration standards covering the expected sample range.
Condition the electrode as specified by its manufacturer.
Use a suitable reference electrode or a combined ISE system.
Measure the potentials of the standards under consistent temperature and stirring conditions.
Construct a calibration relationship between potential and logarithm of ion activity or concentration.
Prepare the unknown sample, controlling pH and ionic strength when required.
Measure the sample potential and determine the analyte result from the calibration relationship.
Check electrode performance with a verification standard.
Equations
| Name | Formula | Meaning |
|---|---|---|
Nernst response for an ion-selective electrode | E = E° + (RT/zF) ln a | For an ideal response written for an ion of charge number z under a suitable sign convention; a is the dimensionless relative activity of the ion. |
Ideal response at 25 °C | E = E° + (0.05916/z) log a | Ideal logarithmic response for the stated convention. The sign and slope depend on the charge of the measured ion and electrode definition. |
Nicolsky–Eisenman relationship | E = E° + (RT/zA F) ln(aA + Σ K(A,B) aB^(zA/zB)) | A simplified form describing the influence of interfering ions; its use requires compatible sign conventions and appropriate selectivity coefficients. |
Units
Potential: volt (V) or millivolt (mV)
Temperature: kelvin (K) or degrees Celsius (°C)
Concentration: mol L⁻¹
Activity: dimensionless
Selectivity coefficient: convention-dependent and generally dimensionless in the stated activity formulation
Examples
A glass membrane electrode measures hydrogen-ion activity and is used for pH measurement.
A fluoride-selective electrode can determine fluoride in water and dental-care samples.
A potassium-selective electrode can measure potassium in biological fluids.
A calcium-selective electrode can be used to investigate calcium levels in aqueous samples.
Applications
Clinical measurement of sodium, potassium, and calcium ions.
Fluoride analysis in drinking water.
Monitoring nitrate and other ions in environmental samples.
Agricultural soil and nutrient analysis.
Industrial monitoring of ionic solutions.
Advantages
Rapid measurements with relatively simple instrumentation.
Small sample volumes may be sufficient.
Suitable for colored or turbid samples when the electrode is compatible.
Can be used for continuous or field measurements.
Many electrodes provide useful selectivity for a target ion.
Limitations
Interfering ions can affect the potential.
Electrode response may depend on pH and ionic strength.
Membranes can deteriorate or become contaminated.
Calibration and regular performance checks are necessary.
The electrode measures activity rather than concentration directly.
Common Mistakes
Treating ion activity and concentration as identical in every solution.
Ignoring interfering ions.
Using standards outside the useful calibration range.
Failing to condition or store the electrode properly.
Assuming that a selective electrode responds exclusively to one ion.
Prerequisites
Electrode potentials and the Nernst equation.
Reference and indicator electrodes.
Concentration and chemical activity.
Basic chemical equilibrium.
Learning Objectives
Explain the construction and working principle of ion-selective electrodes.
Describe the role of selective membranes.
Relate electrode potential to ion activity.
Explain how interfering ions affect selectivity.
Identify common applications of ion-selective electrodes.
Practice Questions
What is an ion-selective electrode?
What is the function of its selective membrane?
How does electrode potential depend on ion activity?
Why is activity used instead of concentration in the ideal Nernst relationship?
What is a selectivity coefficient?
Give three applications of ion-selective electrodes.
Key Takeaway
Ion-selective electrodes measure ionic activity through a membrane-dependent potential response, with accuracy influenced by calibration, sample conditions, and interfering ions.
References
| Title | Type | Url |
|---|---|---|
Harris, Quantitative Chemical Analysis | textbook | |
Bard and Faulkner, Electrochemical Methods: Fundamentals and Applications | textbook | |
IUPAC Gold Book | terminology reference | https://goldbook.iupac.org/ |
Sources and verification
Sources are shown as provided in the project data. Not every source has yet been independently verified. Please report errors or missing citations.
- Source verification pending; check an authoritative textbook, university, standards body, or peer-reviewed source.
Suggest a correction or source · More in Analytical Chemistry