UV–Visible spectrophotometry is based mainly on electronic transitions in atoms, ions, and molecules. When a suitable photon is absorbed, an electron can move from a lower-energy state to a higher-energy state. The instrument measures absorbance or transmittance as a function of wavelength. The resulting spectrum can reveal absorption maxima and characteristic spectral patterns. Quantitative measurements commonly use the Beer–Lambert law or a calibration curve. Samples may be measured in solution, and the solvent, cuvette material, pH, and chemical environment must be selected carefully.
UV–Visible Spectrophotometry
UV–Visible spectrophotometry measures the absorption of ultraviolet and visible radiation by a sample and is widely used for qualitative and quantitative analysis.
Explanation
Variables
| Symbol | Meaning | Unit |
|---|---|---|
A | Absorbance | Dimensionless |
ε | Molar absorption coefficient | L mol⁻¹ cm⁻¹ |
l | Cuvette path length | cm |
c | Analyte concentration | mol L⁻¹ |
λ | Wavelength | nm |
I₀ | Incident intensity | Instrument-dependent |
I | Transmitted intensity | Same as I₀ |
Example or application
A worked example has not yet been added.
Definition
UV–Visible spectrophotometry is an analytical technique that measures the absorption of ultraviolet and visible radiation by a sample as a function of wavelength.
Purpose
To identify absorbing species, examine electronic transitions, and determine the concentration of suitable analytes.
Importance
It is one of the most widely used instrumental methods because many organic molecules, metal complexes, and coloured inorganic species absorb in the UV or visible regions.
Key Concepts
Ultraviolet radiation
Visible radiation
Electronic transitions
Absorption spectrum
Absorption maximum (λmax)
Chromophores
Auxochromes
Beer–Lambert law
Solvent effects
Single-beam and double-beam instruments
Principle
A molecule or ion absorbs UV or visible radiation when the photon energy matches an allowed electronic transition. The measured absorbance at a selected wavelength can be related to analyte concentration under suitable conditions.
Working Process
Select a suitable wavelength range for the analyte.
Prepare the sample and an appropriate solvent blank.
Choose a compatible cuvette, such as quartz for UV measurements.
Establish the blank or reference baseline.
Pass selected wavelengths through the sample.
Measure transmitted radiation and calculate absorbance.
Record absorbance against wavelength to obtain a spectrum.
Use a calibration relationship for quantitative analysis when required.
Equations
| Name | Formula | Description |
|---|---|---|
Beer–Lambert law | A = εlc | Relates absorbance to molar absorption coefficient, path length, and concentration. |
Absorbance | A = log₁₀(I₀/I) | Expresses the attenuation of incident radiation. |
Photon energy | E = hc/λ | Relates the energy of an absorbed photon to its wavelength. |
Units
Wavelength: nm
Absorbance: dimensionless
Concentration: mol L⁻¹
Path length: cm
Molar absorption coefficient: L mol⁻¹ cm⁻¹
Examples
A coloured metal complex can be measured in the visible region.
Benzene and other aromatic compounds show characteristic UV absorption bands.
A pharmaceutical solution can be quantified by measuring absorbance at a validated wavelength.
Applications
Pharmaceutical assay and quality control
Analysis of coloured metal complexes
Determination of nucleic acid and protein concentrations using appropriate methods
Monitoring chemical reactions
Environmental analysis of UV-absorbing pollutants
Study of electronic transitions in molecules
Advantages
Relatively simple and rapid analysis
Often requires limited sample preparation
Suitable for routine quantitative measurements
Can measure a broad range of absorbing compounds
Usually non-destructive under appropriate conditions
Limitations
Compounds without suitable UV–Visible absorption may require derivatization
Overlapping absorption bands can reduce selectivity
Turbid samples can scatter radiation
Solvent and cuvette absorption can interfere
Chemical changes in the sample can alter the spectrum
Common Mistakes
Using glass cuvettes for measurements at wavelengths where glass absorbs strongly
Failing to use a suitable blank
Measuring samples outside the instrument's reliable absorbance range
Ignoring solvent and pH effects
Assuming λmax is always unique to one compound
Prerequisites
Electromagnetic radiation
Absorption and emission
Beer–Lambert law
Basic molecular electronic structure
Learning Objectives
Explain the principle of UV–Visible spectrophotometry
Describe the origin of electronic absorption
Define absorption maximum
Explain the role of blanks and cuvettes
Describe qualitative and quantitative applications
Practice Questions
What is UV–Visible spectrophotometry?
What types of molecular transitions are commonly studied?
Why is quartz preferred for many UV measurements?
What is λmax?
How is UV–Visible spectrophotometry used to determine concentration?
Key Takeaway
UV–Visible spectrophotometry measures electronic absorption and uses characteristic spectra and absorbance relationships to identify and quantify suitable substances.
References
| Title | Type | Url |
|---|---|---|
Skoog, Holler and Crouch, Principles of Instrumental Analysis | textbook | |
Harris, Quantitative Chemical Analysis | textbook | |
IUPAC Gold Book | terminology reference | https://goldbook.iupac.org/ |
Sources and verification
Sources are shown as provided in the project data. Not every source has yet been independently verified. Please report errors or missing citations.
- Source verification pending; check an authoritative textbook, university, standards body, or peer-reviewed source.
Suggest a correction or source · More in Analytical Chemistry