High-performance liquid chromatography is widely used to separate, identify, and quantify compounds in pharmaceutical, environmental, food, and biological samples. A liquid mobile phase is delivered through a column packed with small stationary-phase particles. Because the packed column produces substantial resistance to flow, a high-pressure pump is used to maintain the required mobile-phase flow. The sample is introduced through an injection system, and its components interact differently with the stationary phase as they pass through the column. In reversed-phase HPLC, nonpolar stationary phases such as C18-bonded silica are commonly used with relatively polar mobile phases. Other modes include normal-phase, ion-exchange, and size-exclusion chromatography. Separated components are measured using detectors such as UV-visible, fluorescence, refractive-index, or mass spectrometric detectors. HPLC can operate under isocratic conditions, where mobile-phase composition remains constant, or gradient conditions, where it changes during analysis.
High-Performance Liquid Chromatography (HPLC)
High-performance liquid chromatography is an instrumental separation technique that uses a liquid mobile phase pumped through a packed column under high pressure.
Explanation
Variables
| Symbol | Name | Description |
|---|---|---|
tR | Retention time | Time between sample injection and the appearance of an analyte at the detector. |
tM | Hold-up time | Time required for an unretained compound to pass through the column. |
k | Retention factor | Dimensionless measure of analyte retention relative to the hold-up time. |
α | Selectivity factor | Relative retention of two components. |
Rs | Resolution | Measure of separation between two neighboring peaks. |
F | Mobile-phase flow rate | Volume of mobile phase passing through the system per unit time. |
Example or application
A worked example has not yet been added.
Definition
High-performance liquid chromatography is a liquid chromatographic technique in which a liquid mobile phase is pumped under pressure through a column containing a stationary phase to separate sample components.
Purpose
To separate, identify, and quantitatively determine compounds, including substances that are nonvolatile or thermally unstable.
Importance
HPLC is a major analytical technique in pharmaceutical quality control, clinical research, food testing, environmental monitoring, and chemical laboratories because it can analyze a broad range of compounds with high precision.
Key Concepts
High-pressure pump: Delivers mobile phase through the packed column.
Injector: Introduces a defined sample volume into the mobile-phase stream.
Analytical column: Contains the stationary phase where separation occurs.
Reversed-phase HPLC: Common mode using a nonpolar stationary phase and a relatively polar mobile phase.
Isocratic elution: Mobile-phase composition remains constant.
Gradient elution: Mobile-phase composition changes during the separation.
Retention time: Time required for an analyte to reach the detector.
Chromatogram: Detector response plotted against time.
Principle
Sample components distribute differently between the liquid mobile phase and stationary phase. Their different interactions cause different retention times as the pressurized mobile phase carries them through the column.
Working Process
Prepare and filter the mobile phase when required by the method.
Degas the mobile phase or use an appropriate solvent-degassing system.
Pump the mobile phase through the chromatographic system at a controlled flow rate.
Equilibrate the column with the selected mobile-phase composition.
Inject a measured sample volume into the flowing mobile phase.
Allow sample components to interact with the stationary phase and separate.
Pass the separated components through a suitable detector.
Record the chromatogram and use calibrated responses for quantitative analysis.
Equations
Retention factor: k = (tR - tM) / tM
Selectivity factor: α = k2 / k1, where k2 > k1
Resolution: Rs = 2(tR2 - tR1) / (w1 + w2)
Units
Mobile-phase flow rate: millilitre per minute (mL min⁻¹)
Pressure: pascal (Pa), bar, or megapascal (MPa)
Retention time: minute (min)
Injection volume: microlitre (µL)
Column temperature: degree Celsius (°C)
Retention factor: dimensionless
Resolution: dimensionless
Examples
Quantifying an active pharmaceutical ingredient in a tablet.
Separating caffeine and other compounds in a beverage sample.
Analyzing preservatives in food products.
Measuring selected environmental contaminants in water.
Applications
Pharmaceutical assay and impurity profiling.
Food and beverage analysis.
Environmental contaminant measurement.
Clinical and biochemical analysis.
Purification and characterization of chemical compounds.
Advantages
Suitable for many nonvolatile and thermally unstable compounds.
Provides high separation efficiency with appropriate columns.
Can be automated for routine analysis.
Supports both qualitative and quantitative analysis.
Can be coupled with detectors such as mass spectrometers.
Limitations
Requires specialized equipment and regular maintenance.
Consumes organic solvents in many methods.
Column performance can deteriorate with contamination or unsuitable samples.
Method development may be time-consuming.
High pressure requires compatible tubing, fittings, and components.
Common Mistakes
Confusing isocratic and gradient elution.
Injecting samples containing particles that can block the system.
Using incompatible solvents or mobile-phase additives.
Ignoring column equilibration.
Assuming retention time alone proves compound identity.
Failing to consider detector compatibility with the mobile phase.
Prerequisites
Chromatographic separation principles.
Stationary and mobile phases.
Retention time and resolution.
Solutions, solvents, and basic fluid-flow concepts.
Learning Objectives
Define HPLC and explain its operating principle.
Identify the major components of an HPLC system.
Distinguish isocratic and gradient elution.
Explain the role of the packed column and high-pressure pump.
Recognize major HPLC applications and limitations.
Practice Questions
What is HPLC?
Why does HPLC require a high-pressure pump?
Differentiate isocratic and gradient elution.
Why is reversed-phase HPLC widely used?
Give four applications of HPLC in analytical chemistry.
Key Takeaway
HPLC uses a pressurized liquid mobile phase and a packed stationary-phase column to separate compounds that may not be suitable for gas chromatography.
References
| Title | Type | Url |
|---|---|---|
Snyder, Kirkland and Dolan, Introduction to Modern Liquid Chromatography | textbook | |
Skoog, Holler and Crouch, Principles of Instrumental Analysis | textbook | |
IUPAC Gold Book | terminology reference | https://goldbook.iupac.org/ |
Sources and verification
Sources are shown as provided in the project data. Not every source has yet been independently verified. Please report errors or missing citations.
- Source verification pending; check an authoritative textbook, university, standards body, or peer-reviewed source.
Suggest a correction or source · More in Analytical Chemistry