Electron ionization (EI) and chemical ionization (CI) are important gas-phase ionization methods. In electron ionization, gaseous sample molecules are exposed to energetic electrons, commonly with an energy of 70 eV in conventional instruments. An electron is removed from the molecule, producing a molecular radical cation when that ion is formed. The excess internal energy often causes extensive fragmentation, creating a characteristic pattern that can help identify the compound. In chemical ionization, a reagent gas is first ionized. The reagent-gas ions then react with sample molecules, commonly through proton transfer, to form analyte ions. Chemical ionization is generally softer than electron ionization and often produces a prominent protonated molecule. Both methods are widely associated with gas chromatography–mass spectrometry.
Electron Ionization and Chemical Ionization
Electron ionization and chemical ionization are gas-phase ionization techniques commonly used in mass spectrometry, particularly for volatile compounds.
Explanation
Variables
| Symbol | Name | Description |
|---|---|---|
M | Analyte molecule | Gaseous sample molecule undergoing ionization. |
e⁻ | Electron | Electron used to ionize molecules in electron ionization. |
R | Reagent gas species | Species derived from the reagent gas used in chemical ionization. |
RH⁺ | Protonated reagent ion | Reagent-gas ion capable of transferring a proton to the analyte. |
M⁺• | Molecular radical cation | Positively charged molecular ion formed by electron removal. |
Example or application
A worked example has not yet been added.
Definition
Electron ionization is a gas-phase technique in which energetic electrons remove an electron from a sample molecule, whereas chemical ionization uses reagent-gas ions to ionize analyte molecules through gas-phase chemical reactions.
Purpose
To produce gas-phase ions from suitable volatile compounds for molecular-mass determination and structural identification.
Importance
EI provides reproducible fragmentation patterns useful for spectral-library matching, while CI often provides clearer molecular-mass-related information through reduced fragmentation.
Key Concepts
Electron ionization (EI): Ion formation through electron impact and electron removal.
Molecular radical cation: A positively charged molecular ion containing an unpaired electron.
Fragmentation: Breakdown of an energized molecular ion into smaller ions.
Reagent gas: Gas ionized first during chemical ionization.
Chemical ionization (CI): Ionization of analyte molecules through reactions with reagent-gas ions.
Proton transfer: Transfer of a proton from a reagent-gas ion to an analyte.
Hard ionization: EI is generally considered a hard ionization method.
Soft ionization: CI generally produces less fragmentation than EI under comparable analytical conditions.
Principle
In EI, energetic electrons collide with gaseous molecules and can remove electrons, forming radical cations. In CI, reagent-gas ions react with analyte molecules, often transferring a proton to form protonated analyte ions.
Working Process
The sample is vaporized and introduced into the ion source.
For EI, an electron beam interacts directly with the gaseous sample molecules.
For CI, a reagent gas is introduced and ionized before significant analyte ion formation.
EI produces molecular radical cations and fragment ions; CI produces analyte ions through reagent-ion reactions.
The resulting ions are extracted into the mass analyzer.
The mass spectrum is recorded and interpreted using molecular-related and fragment-ion peaks.
Equations
M + e⁻ → M⁺• + 2e⁻
R + e⁻ → R⁺• + 2e⁻
RH⁺ + M → [M + H]⁺ + R
m/z = m / z
Units
Electron energy: electronvolt (eV)
Ion mass: dalton (Da)
Mass-to-charge ratio: m/z
Source pressure: pascal (Pa) or millibar (mbar)
Gas flow: millilitre per minute (mL min⁻¹), where applicable
Examples
EI is commonly used to obtain fragmentation patterns of volatile organic compounds.
An EI spectrum may contain a molecular radical cation and several fragment-ion peaks.
Positive chemical ionization may produce a prominent [M + H]⁺ ion.
Methane, isobutane, and ammonia are examples of gases that may be used as chemical-ionization reagents, depending on the method.
Applications
Identification of volatile organic compounds.
GC-MS analysis of environmental pollutants.
Forensic analysis of chemical residues.
Analysis of pharmaceutical compounds suitable for gas-phase ionization.
Comparison of EI spectra with reference spectral libraries.
Confirmation of molecular mass using chemical ionization.
Advantages
EI produces reproducible fragmentation patterns useful for library searching.
EI is well established for many volatile and thermally stable compounds.
CI often produces less fragmentation than EI.
CI can provide useful molecular-mass-related peaks.
Both techniques can be integrated with gas chromatography.
Limitations
Both methods generally require analytes that can enter the gas phase.
Thermally unstable compounds may decompose during vaporization.
EI fragmentation can make the molecular ion weak or absent.
CI spectra depend on the reagent gas and reaction conditions.
Neither method is universally suitable for large, nonvolatile biomolecules.
Common Mistakes
Assuming EI always produces a strong molecular ion peak.
Treating CI as a technique that directly bombards analytes with electrons in the same way as EI.
Assuming the peak at [M + H]⁺ has the same m/z as the neutral molecular mass.
Ignoring the influence of reagent-gas selection in CI.
Assuming that all compounds are sufficiently volatile and thermally stable for these methods.
Prerequisites
Introduction to mass spectrometry.
Gas-phase molecular behavior.
Electron removal and radical ions.
Proton-transfer reactions.
Learning Objectives
Describe the mechanism of electron ionization.
Explain how chemical ionization uses reagent-gas ions.
Distinguish molecular radical cations from protonated molecules.
Compare the fragmentation behavior of EI and CI.
Identify common applications and limitations of both techniques.
Practice Questions
What happens when a molecule undergoes electron ionization?
What is a molecular radical cation?
Why is EI generally classified as a hard ionization method?
How does chemical ionization differ from electron ionization?
What is the role of a reagent gas in CI?
Why can CI be useful when the EI molecular ion is weak?
Key Takeaway
EI generates ions by direct electron impact and often produces extensive fragmentation, whereas CI uses reagent-ion reactions and generally produces less fragmentation.
References
| Title | Type | Url |
|---|---|---|
Gross, Mass Spectrometry: A Textbook | textbook | |
Hoffmann and Stroobant, Mass Spectrometry: Principles and Applications | textbook | |
Skoog, Holler and Crouch, Principles of Instrumental Analysis | textbook | |
IUPAC Gold Book | terminology reference | https://goldbook.iupac.org/ |
Sources and verification
Sources are shown as provided in the project data. Not every source has yet been independently verified. Please report errors or missing citations.
- Source verification pending; check an authoritative textbook, university, standards body, or peer-reviewed source.
Suggest a correction or source · More in Analytical Chemistry