Electrospray ionization and matrix-assisted laser desorption/ionization are widely used for the mass spectrometric analysis of biomolecules and other compounds that are difficult to vaporize without decomposition. In ESI, a solution containing the analyte passes through a narrow capillary held at a high electrical potential. Charged droplets form at the capillary tip. As solvent evaporates, the droplets shrink and eventually release gas-phase ions through accepted ion-formation mechanisms. ESI commonly produces multiply charged ions, which allows large molecules to be measured within the m/z range of many mass analyzers. In MALDI, the sample is mixed with a suitable light-absorbing matrix and deposited on a target. Laser energy is absorbed by the matrix, helping desorb and ionize the analyte. MALDI commonly produces singly charged ions. Both methods are considered soft ionization techniques, although fragmentation can still occur under suitable conditions.
Electrospray Ionization and MALDI
Electrospray ionization (ESI) and matrix-assisted laser desorption/ionization (MALDI) are soft ionization techniques used to analyze large, polar, and thermally sensitive molecules.
Explanation
Variables
| Symbol | Name | Description |
|---|---|---|
M | Analyte molecule | Molecule being analyzed by ESI or MALDI. |
z | Charge number | Number of elementary charges carried by an ion. |
n | Number of transferred protons | Number of protons added or removed in the simplified ion-formation equations. |
m | Ion mass | Mass of the resulting charged analyte species. |
m/z | Mass-to-charge ratio | Value measured by the mass analyzer. |
Example or application
A worked example has not yet been added.
Definition
Electrospray ionization is a soft ionization method that generates gas-phase ions from charged solution droplets, while MALDI is a soft ionization method that uses a laser-absorbing matrix to assist the desorption and ionization of sample molecules.
Purpose
To analyze large, polar, nonvolatile, or thermally sensitive molecules that are difficult to examine using conventional gas-phase ionization methods.
Importance
ESI and MALDI have greatly expanded mass spectrometry applications in proteomics, peptide analysis, pharmaceutical research, polymer characterization, and biomolecular studies.
Key Concepts
Electrospray ionization (ESI): Formation of gas-phase ions from charged droplets produced from a solution.
Charged droplets: Small liquid droplets carrying excess electrical charge.
Desolvation: Removal of solvent from droplets or analyte-containing particles.
Multiply charged ions: Ions carrying several elementary charges, commonly observed in ESI.
Matrix-assisted laser desorption/ionization (MALDI): Laser-assisted ionization using an energy-absorbing matrix.
Matrix: A compound that absorbs laser energy and assists analyte desorption and ionization.
Singly charged ions: Ions carrying one elementary charge, commonly produced by MALDI.
Soft ionization: Ionization that generally preserves molecular-related ions with limited fragmentation.
Principle
ESI uses an electric field to create charged droplets from a sample solution. Solvent evaporation and droplet evolution lead to gas-phase ions. MALDI uses a matrix that absorbs laser energy and assists the release and ionization of analyte molecules into the gas phase.
Working Process
ESI sample preparation: The analyte is dissolved in a suitable solvent or solvent mixture.
ESI droplet formation: A high electrical potential at the capillary produces charged droplets.
ESI ion release: Solvent evaporation and droplet evolution lead to gas-phase ions.
MALDI sample preparation: The analyte is mixed with a suitable matrix and deposited on a target plate.
MALDI laser interaction: A laser pulse is absorbed by the matrix.
MALDI ion formation: Analyte molecules are desorbed and ionized.
Mass analysis: Ions from either source enter a mass analyzer for separation and detection.
Equations
m/z = m / z
m = z × (m/z)
ESI protonation: M + nH⁺ → [M + nH]ⁿ⁺
ESI deprotonation: M - nH⁺ → [M - nH]ⁿ⁻
Units
Ion mass: dalton (Da) or kilodalton (kDa)
Mass-to-charge ratio: m/z
Electrical potential: volt (V)
Laser wavelength: nanometre (nm)
Sample concentration: mg mL⁻¹ or µg mL⁻¹, depending on the method
Examples
ESI can generate multiply charged ions from proteins and peptides.
ESI is commonly coupled with liquid chromatography for pharmaceutical and biomolecular analysis.
MALDI is widely used to analyze proteins, peptides, polymers, and microbial samples.
MALDI commonly produces singly charged molecular-related ions, simplifying some spectra.
Applications
Protein and peptide identification.
Proteomics and biomolecular characterization.
Pharmaceutical compound analysis.
Polymer molecular-mass characterization.
LC-MS analysis of polar and thermally sensitive compounds.
Analysis of large biomolecules and complex biological samples.
Advantages
Both techniques can analyze compounds that are difficult to vaporize directly.
ESI can be coupled readily with liquid chromatography.
ESI can generate multiply charged ions from large molecules.
MALDI commonly produces relatively simple spectra with predominantly singly charged ions.
Both methods are useful for large and thermally sensitive molecules.
Limitations
ESI can be affected by salts, detergents, and matrix-related ion suppression.
ESI may produce complex distributions of charge states and adducts.
MALDI results depend on matrix selection and sample crystallization.
MALDI can show spot-to-spot variation and matrix-related background signals.
Both techniques require suitable sample preparation and optimized operating conditions.
Common Mistakes
Assuming ESI normally produces only singly charged ions.
Assuming MALDI and ESI use the same ion-formation mechanism.
Confusing the matrix used in MALDI with the analyte itself.
Assuming soft ionization means fragmentation can never occur.
Ignoring salt and solvent effects in ESI.
Prerequisites
Basic principles of mass spectrometry.
Ionization techniques.
Acid-base chemistry and proton transfer.
Basic solution chemistry.
Learning Objectives
Explain the formation of charged droplets in ESI.
Describe the role of solvent evaporation in ESI ion production.
Explain the role of the matrix and laser in MALDI.
Compare the charge-state distributions commonly produced by ESI and MALDI.
Identify suitable applications and limitations of both methods.
Practice Questions
What is electrospray ionization?
How are gas-phase ions produced in ESI?
Why can ESI generate multiply charged ions?
What is the role of the matrix in MALDI?
How does MALDI differ from ESI in its ion-generation process?
Why are ESI and MALDI useful for biomolecular analysis?
Key Takeaway
ESI produces ions from charged solution droplets and often generates multiple charge states, whereas MALDI uses a laser-absorbing matrix and commonly produces singly charged ions.
References
| Title | Type | Url |
|---|---|---|
Gross, Mass Spectrometry: A Textbook | textbook | |
Hoffmann and Stroobant, Mass Spectrometry: Principles and Applications | textbook | |
Cole, Electrospray Ionization Mass Spectrometry | textbook | |
IUPAC Gold Book | terminology reference | https://goldbook.iupac.org/ |