Instrument characteristics are commonly divided into static and dynamic characteristics. Static characteristics describe instrument behavior when the measurand is constant or changes sufficiently slowly. Dynamic characteristics describe how the instrument responds when the measurand changes with time.
Accuracy describes closeness of agreement between a measured value and a reference value. Precision describes agreement among repeated measurements under specified conditions. Neither term should be confused with resolution. A highly precise instrument can still have a systematic bias, and a high-resolution display does not necessarily indicate high accuracy.
Other important characteristics include sensitivity, linearity, hysteresis, repeatability, reproducibility, stability, drift, dead band, threshold, range, response time and loading effect. These characteristics are often specified by manufacturers, but their actual values may depend on calibration, installation, environmental conditions and measurement procedure.
A complete instrument specification should be interpreted in context. For example, an accuracy specification expressed as a percentage of full scale differs from one expressed as a percentage of reading. A response-time specification may also depend on the sensor, protective sheath, signal filtering and measurement environment.
Measuring Instrument Characteristics and Performance Parameters
Instrument characteristics describe how a measuring instrument responds to an input, how closely its indications agree with reference values, how consistently it performs and how its behavior changes with time and operating conditions. Understanding these characteristics is necessary for selecting instruments, evaluating measurement quality and interpreting results.
Absolute Error
e = x_m - x_ref
Relative Error
e_r = (x_m - x_ref) / x_ref
Percentage Error
e_% = |x_m - x_ref| / |x_ref| × 100%
Sensitivity
S = Δy / Δx
Linearity Deviation
L = max|y_actual - y_reference|
Note
These expressions are simplified evaluation measures. A complete uncertainty evaluation must consider the measurement model, uncertainty sources and relevant probability distributions.
Explanation
Variables
| Symbol | Meaning | Unit |
|---|---|---|
x_m | Measured or indicated value | Unit of the measurand |
x_ref | Reference value | Unit of the measurand |
e | Signed measurement error | Unit of the measurand |
e_r | Relative error | Dimensionless |
S | Sensitivity | Output unit per input unit |
Δx | Change in input | Unit of the measurand |
Δy | Corresponding change in output | Output unit |
Example or application
A worked example has not yet been added.
Definition
Accuracy
Closeness of agreement between a measured quantity value and a true quantity value or an accepted reference value.
Precision
Closeness of agreement among measured values obtained by replicate measurements under specified conditions.
Repeatability
Precision under repeatability conditions, such as the same method, operator, instrument, location and short time interval.
Reproducibility
Precision under changed conditions, such as different laboratories, operators or instruments.
Linearity
Closeness of an instrument's input-output relationship to a specified straight-line relationship.
Hysteresis
Dependence of an instrument's output on the direction or history of input change.
Stability
Ability of an instrument to maintain its metrological properties over time.
Drift
A continuous or incremental change in an indication over time that is not caused by a corresponding change in the measurand.
Loading Effect
The disturbance introduced by a measuring instrument into the system being measured.
Purpose
To evaluate instrument suitability, compare alternative instruments, identify limitations, establish calibration requirements and determine whether an instrument can meet a measurement task's uncertainty and performance requirements.
Importance
Instrument performance directly influences the reliability of experimental data, product tolerances, process control and safety decisions. Understanding performance parameters prevents the misuse of instruments whose range, accuracy, response or stability is unsuitable.
Static Characteristics
| Characteristic | Explanation |
|---|---|
Accuracy | Indicates closeness of agreement with a reference. Accuracy is a qualitative concept and should not be treated as a numerical property without a stated specification or uncertainty evaluation. |
Precision | Describes agreement among repeated results. It is affected by random variation and measurement conditions. |
Sensitivity | Describes how much the output changes for a given change in input. |
Linearity | Describes how closely the actual calibration curve follows a specified linear relationship. |
Hysteresis | Describes different indications for the same input depending on whether the input is approached from above or below. |
Repeatability | Describes consistency under the same measurement conditions over a short period. |
Reproducibility | Describes consistency when specified conditions, such as operators or laboratories, change. |
Stability and drift | Describe the ability to maintain performance over time and the change in indication that may occur without a corresponding measurand change. |
Dead band | An input interval over which a change in input produces no detectable output change. |
Threshold | The smallest input change or input level that produces a detectable response under specified conditions. |
Dynamic Characteristics
| Characteristic | Explanation |
|---|---|
Response time | Time required for an instrument to respond to a change in the measurand, often specified using a defined percentage of the final response. |
Time constant | For a first-order system, the time required to reach approximately 63.2% of the total response to a step input. |
Rise time | Time taken for an output to rise between specified percentages of its final value. |
Settling time | Time required for the output to enter and remain within a specified band around its final value. |
Frequency response | Describes how an instrument's gain and phase response vary with the frequency of a changing input. |
Performance Specifications
Range
The interval between specified lower and upper limits of measurement.
Span
The difference between the upper and lower range limits.
Resolution
The smallest change in the measured quantity that produces a perceptible change in indication.
Sensitivity
The ratio of output change to input change.
Accuracy Specification
May be stated as a percentage of reading, full scale, span or a fixed quantity, depending on the manufacturer.
Repeatability Specification
May be expressed as a standard deviation, range or other statistical measure under specified conditions.
Environmental Rating
Describes permitted operating conditions such as temperature, humidity, vibration and ingress protection.
Worked Examples
| Problem | Solution |
|---|---|
A reference instrument indicates 100.0 mm, while a test instrument indicates 100.8 mm. Determine the signed error and absolute percentage error. | Signed error = 100.8 mm − 100.0 mm = +0.8 mm. Absolute percentage error = (0.8/100.0) × 100% = 0.8%. This comparison does not by itself establish the full uncertainty of either instrument. |
An instrument output changes by 4.0 V when its input changes by 2.0 units of the measurand. Find its average sensitivity over this interval. | S = Δy/Δx = 4.0 V/2.0 units = 2.0 V per input unit. |
A first-order sensor has a time constant of 3 s. Approximately how long does it take to reach 63.2% of its final response after a step input? | For a first-order system, the time constant is the time to reach approximately 63.2% of the final response. Therefore, the required time is 3 s. |
Measurement Quality
Systematic Effects
Effects that cause a consistent or predictable deviation, such as calibration bias, zero offset or temperature-dependent error.
Random Effects
Effects that produce unpredictable variation among repeated measurements.
Uncertainty
A non-negative parameter characterizing the dispersion of quantity values attributed to a measurand, based on the information used.
Calibration
An operation that establishes a relation between reference quantity values and instrument indications, including their uncertainties.
Correction
A value applied to compensate for an estimated systematic effect.
Common Mistakes
Using accuracy and precision as interchangeable terms.
Assuming high resolution means high accuracy.
Ignoring hysteresis when measuring a changing input.
Comparing accuracy specifications that use different reference bases.
Ignoring response time when measuring rapidly changing quantities.
Assuming calibration permanently removes all measurement errors.
Applications
Selecting laboratory instruments
Industrial process measurement
Quality control and dimensional inspection
Sensor characterization
Calibration laboratories
Scientific experiments and data acquisition
Reference
JCGM 100, Evaluation of Measurement Data — Guide to the Expression of Uncertainty in Measurement (GUM).
JCGM 200, International Vocabulary of Metrology (VIM).
NIST, Special Publication 811, Guide for the Use of the International System of Units (SI).
Related Topic
Introduction to Measuring Instruments and Instrument Classification
Resolution, Sensitivity, Range and Threshold of Instruments
Measurement Error and Uncertainty
Calibration and Traceability
Sources and verification
Sources are shown as provided in the project data. Not every source has yet been independently verified. Please report errors or missing citations.
- Source verification pending; check an authoritative textbook, university, standards body, or peer-reviewed source.
Suggest a correction or source · More in Units & Measurement