A measurement result is produced by a complete measurement system rather than by an instrument alone. The result depends on the measurand, measurement principle, instrument, calibration, operator, environment, sample condition, data acquisition and mathematical model. A weakness in any of these components can influence the final result.
Instrument-related errors arise from the construction, adjustment, calibration and operating characteristics of measuring equipment. Examples include zero offset, scale-factor error, nonlinearity, hysteresis, backlash, friction, resolution limitations, electrical noise, drift and sensor ageing. An instrument may also respond differently under different loads, frequencies, temperatures or measurement ranges.
Environmental errors arise when external conditions influence the measuring instrument, sample or measurement process. Temperature can cause thermal expansion, alter electrical resistance or change sensor sensitivity. Humidity can affect materials and insulation. Vibration can disturb dimensional measurements, while electromagnetic interference can affect electrical signals. Pressure, airflow, lighting and acoustic noise can also influence particular measurement methods.
Observational errors arise from how a person reads, positions, operates or interprets a measuring system. Parallax occurs when an analogue scale is viewed from an unsuitable angle. Incorrect alignment can affect length and angle measurements. Reaction time can affect manual timing. Inconsistent contact force can influence measurements made using mechanical instruments.
Method-related errors occur when the chosen procedure does not correctly represent the measurand or when necessary corrections are omitted. Measuring the diameter of a non-circular object at only one location may not represent the intended dimension. Measuring a moving object with a method that assumes it is stationary can also produce a method-dependent result. The measurement procedure must define the quantity sufficiently clearly.
Sample-related effects occur when the object or material changes during measurement. Examples include thermal expansion, evaporation, moisture absorption, deformation under contact pressure, chemical reaction, contamination and ageing. In biological or environmental measurements, the sample may also vary naturally with location or time.
Reference and calibration-related effects arise from the uncertainty or limitations of reference standards, calibration procedures, reference materials and traceability chains. A calibration result is not automatically exact. The reference, calibration method, environmental conditions and instrument stability all contribute to the final measurement uncertainty.
Data-processing errors arise during conversion, calculation, software processing, filtering, rounding, storage or transmission. An incorrect unit, formula, calibration coefficient or software setting can systematically affect a large set of results. A transcription mistake can affect an individual result. Automated processing therefore requires validation, version control and independent checks appropriate to the risk.
The influence of a source depends on the measurement task. Temperature may be negligible in a rough classroom measurement but significant in precision dimensional metrology. Reaction time may be important when timing a short event manually but negligible when measuring a long duration with an automated clock. Error investigation should therefore be based on the actual measurement model and required performance.
Sources and Mechanisms of Measurement Errors
Measurement errors arise from interactions among instruments, measurement methods, operators, environmental conditions, reference standards, samples and data-processing procedures. Understanding these sources makes it possible to identify error mechanisms, determine which effects influence a result and design suitable controls. This topic examines common physical, instrumental, environmental, observational and computational causes of measurement deviations.
General measurement model: Y = f(X₁, X₂, ..., Xₙ)
For small input variations, the first-order sensitivity coefficient is cᵢ = ∂f/∂Xᵢ.
A simplified independent-input uncertainty model is u_c²(Y) = Σ[cᵢ²u²(Xᵢ)].
For correlated inputs, covariance terms must also be included:
u_c²(Y) = Σ[cᵢ²u²(Xᵢ)] + 2Σᵢ<ⱼ[cᵢcⱼu(Xᵢ, Xⱼ)].
Explanation
Variables
| Symbol | Meaning | Unit |
|---|---|---|
Y | Output quantity or measurement result | Unit of the output quantity |
Xᵢ | ith input quantity in the measurement model | Unit appropriate to that input quantity |
f | Mathematical measurement model relating inputs to the output | Depends on the model |
cᵢ | Sensitivity coefficient showing how the output responds to a change in input Xᵢ | Output unit divided by input unit |
u(Xᵢ) | Standard uncertainty associated with input quantity Xᵢ | Same unit as Xᵢ |
u_c(Y) | Combined standard uncertainty associated with output quantity Y | Same unit as Y |
u(Xᵢ, Xⱼ) | Covariance between input quantities Xᵢ and Xⱼ | Product of the units of Xᵢ and Xⱼ |
Example or application
A worked example has not yet been added.
Definition
A source of measurement error is an instrument, environmental, procedural, observational, sample-related, reference-related or data-processing influence that contributes to the difference between a measured quantity value and a reference quantity value.
Purpose
The purpose of studying measurement-error sources and mechanisms is to understand how the measurement system produces deviations, identify influential factors and determine where controls, corrections, calibration or improved procedures are needed.
Importance
Identifying sources is necessary for a meaningful uncertainty evaluation and for improving measurement reliability. Without understanding the mechanism, an investigator may apply an ineffective correction, overlook an important influence or incorrectly attribute a systematic effect to random variation.
Classification Of Sources
| Source | Mechanisms | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Instrument-related |
| ||||||||||||||||
Environmental |
| ||||||||||||||||
Operator and observation |
| ||||||||||||||||
Measurement method |
| ||||||||||||||||
Sample and measurand |
| ||||||||||||||||
Reference and calibration |
| ||||||||||||||||
Data processing and software |
|
Mechanism Investigation
Steps
Define the measurand and the conditions under which it is measured.
Draw or describe the measurement system, including instruments, references, samples and data-processing stages.
Identify possible input quantities and physical influences.
Determine how each influence can affect the measurement result.
Estimate the sensitivity of the result to important inputs.
Use calibration, experiments, environmental monitoring or comparison methods to investigate influential effects.
Apply justified corrections where appropriate and evaluate the uncertainty remaining after correction.
Document assumptions, procedures, instrument status and environmental conditions.
Measurement Model Example
Quantity
Density of a uniform sample
Model
ρ = m/V
Potential Sources
Mass indication and balance calibration
Volume measurement and dimensional geometry
Temperature-dependent sample expansion
Air buoyancy where relevant to the weighing method
Sample nonuniformity and trapped air
Unit conversion and numerical processing
Explanation
The density result depends on both mass and volume. An accurate mass measurement alone does not guarantee an accurate density if the volume measurement or sample model is inadequate.
Worked Examples
| Title | Problem | Solution | Interpretation |
|---|---|---|---|
Temperature influence on a length measurement | A metal component is measured at a temperature different from its reference temperature. Explain why this can influence the result. | For a uniform material under a small temperature change, the approximate length change is ΔL = αL₀ΔT, where α is the coefficient of linear thermal expansion, L₀ is the reference length and ΔT is the temperature difference. | The measured length can differ because the sample itself expands or contracts. The effect should be considered when the required measurement accuracy makes it significant. |
Identifying a data-processing source | A speed is calculated from a distance of 100 m and a time of 20 s, but the software uses 10 s. | Correct speed = 100 m / 20 s = 5 m/s. Incorrect software result = 100 m / 10 s = 10 m/s. | The error originates in the input data or processing configuration rather than necessarily in the measuring instruments. |
Sensitivity to input uncertainty | A quantity is calculated using Y = 2X. Explain how a small uncertainty in X affects Y. | The sensitivity coefficient is ∂Y/∂X = 2. For a small input uncertainty u(X), the corresponding standard uncertainty contribution is approximately 2u(X). | A measurement model can amplify or reduce the influence of input uncertainty. |
Common Misconceptions
All measurement errors do not originate in the instrument.
An environmental influence may affect the sample, instrument or both.
Calibration does not remove every possible measurement influence.
A high-resolution instrument can still have substantial calibration or environmental errors.
A correct calculation cannot compensate for an incorrectly defined measurand.
Software-generated results are not automatically free from measurement errors.
A correction should be supported by evidence and should not be applied without considering its uncertainty.
Applications
Laboratory measurement-system design
Dimensional metrology and precision manufacturing
Environmental monitoring
Sensor development and testing
Electrical and electronic measurements
Mass, force and pressure measurement
Scientific data processing
Calibration and quality-assurance laboratories
Related Topics
Introduction to Measurement Errors
Systematic, Random and Gross Errors
Measurement Instrument Characteristics
Measurement Uncertainty
Calibration and Metrological Traceability
Reference
| Title | Organization | Url |
|---|---|---|
Guide to the Expression of Uncertainty in Measurement | JCGM | https://www.bipm.org/en/committees/jc/jcgm/publications |
International Vocabulary of Metrology (VIM), JCGM 200 | Joint Committee for Guides in Metrology | https://www.bipm.org/en/committees/jc/jcgm/publications |
NIST/SEMATECH e-Handbook of Statistical Methods | National Institute of Standards and Technology | https://www.itl.nist.gov/div898/handbook/ |
Related Topic
SCI-REF-003-073
SCI-REF-003-074
SCI-REF-003-076
SCI-REF-003-080
Sources and verification
Sources are shown as provided in the project data. Not every source has yet been independently verified. Please report errors or missing citations.