✦ScienexySCIENCE KNOWLEDGE LIBRARY
reference · SCI-REF-003-093

Calibration Principles, Procedures and Calibration Certificates

Calibration establishes a documented relationship between reference quantity values and the indications of a measuring instrument or measurement system under specified conditions. It provides information about instrument behavior, corrections, uncertainty and metrological traceability, supporting reliable use of measurement equipment.

Indication error: E = I − R. Correction: C = R − I = −E, where I is the instrument indication and R is the reference value. A corrected result may be estimated as I_corrected = I + C.

Explanation

Calibration is a fundamental activity in metrology. It determines how an instrument's indications relate to reference values under defined conditions. During calibration, an instrument is compared with a suitable reference standard or measurement system. Several reference values may be applied across the instrument's measurement range to identify offset, sensitivity, nonlinearity and other characteristics. Calibration does not necessarily involve adjusting the instrument. Calibration establishes the relationship between indications and reference values, whereas adjustment changes the instrument so that its indications correspond more closely to desired values. Verification determines whether specified requirements are fulfilled. These activities are related but have different purposes. A calibration procedure should define the measurand, instrument identification, reference standards, environmental conditions, measurement points, repetitions, calculation method, uncertainty evaluation and acceptance criteria where applicable. The calibration result may be expressed through correction values, calibration curves, indication errors or other suitable relationships. A calibration certificate records the results and supporting information. It should identify the instrument, calibration date, method, reference standards, measurement conditions, results and uncertainty. A statement of conformity should only be included when a suitable specification and decision rule have been established. Calibration intervals should be selected according to instrument stability, usage, environmental conditions, historical performance and risk. A fixed annual interval is not automatically appropriate for every instrument.

Variables

SymbolMeaningSi Unit

I

Indication provided by the instrument

Unit of the measured quantity

R

Reference quantity value

Same unit as I

E

Indication error

Same unit as I

C

Calibration correction

Same unit as I

u_c

Combined standard uncertainty associated with the calibration result

Same unit as the calibrated quantity

Example or application

A worked example has not yet been added.

Definition

Calibration is an operation that, under specified conditions, in a first step establishes a relation between quantity values with measurement uncertainties provided by measurement standards and corresponding indications with associated measurement uncertainties, and in a second step uses this information to establish a relation for obtaining a measurement result from an indication.

Purpose

To establish instrument performance relative to reference standards, determine corrections and uncertainty, and provide documented evidence supporting reliable measurement.

Importance

Calibration is essential for measurement traceability, instrument reliability, quality control, regulatory compliance, laboratory competence and confidence in scientific and industrial results.

Calibration Principles

PrincipleDescription

Reference comparison

The instrument is compared with a suitable reference standard or measurement system.

Defined conditions

Calibration conditions must be specified because temperature, humidity, pressure, orientation and other influences may affect results.

Measurement range

Calibration points should represent the range and operating conditions for which the instrument is intended to be used.

Uncertainty evaluation

Reference uncertainty, repeatability, resolution, environmental influences and other relevant contributions are evaluated.

Traceability

Reference values should be connected to suitable measurement standards through a documented calibration chain where traceability is required.

Documented results

Calibration data, corrections, uncertainty and conditions should be recorded.

Calibration Adjustment And Verification

ActivityMeaningOutput

Calibration

Establishes the relationship between reference values and instrument indications.

Calibration results, corrections and associated uncertainty.

Adjustment

Changes the instrument so that its indications correspond to prescribed values.

An adjusted instrument requiring appropriate verification or recalibration.

Verification

Provides objective evidence that specified requirements have been fulfilled.

A conformity decision against defined requirements.

Calibration Procedure

  • Identify the instrument and its intended measurement range.

  • Review previous calibration records and instrument condition.

  • Select suitable reference standards with appropriate uncertainty.

  • Define calibration points and the calibration method.

  • Control relevant environmental conditions.

  • Allow equipment to stabilize where necessary.

  • Perform reference and instrument readings.

  • Repeat readings where required to assess variation.

  • Calculate indication errors and calibration corrections.

  • Evaluate calibration uncertainty.

  • Compare results with acceptance requirements if a decision is required.

  • Prepare the calibration report or certificate.

  • Label and return the instrument with its calibration status clearly communicated.

Calibration Methods

MethodDescriptionExample

Direct comparison

Instrument indications are compared directly with values provided by a reference standard.

Comparing a thermometer with a calibrated reference thermometer.

Substitution

An unknown quantity is replaced by a reference quantity under comparable measurement conditions.

Replacing an unknown mass with a calibrated reference mass.

Differential calibration

A small difference between the instrument under calibration and a reference is measured.

Comparing two closely matched electrical voltage sources.

Indirect calibration

Calibration is performed through a measurement model involving several quantities.

Calibrating a flow measurement system using measured mass and time.

Calibration Certificate Contents

  • Unique certificate identification.

  • Name and address of the calibration laboratory.

  • Customer and instrument identification.

  • Instrument manufacturer, model and serial number where applicable.

  • Calibration date and relevant dates of receipt or testing.

  • Calibration method or procedure.

  • Environmental and operational conditions where relevant.

  • Reference standards and relevant traceability information.

  • Calibration results and measurement units.

  • Measurement uncertainty and coverage information where applicable.

  • Corrections or calibration curves when relevant.

  • Conformity statements and decision rules when requested and justified.

  • Authorized approval and other required certificate information.

Worked Example

Problem

A thermometer indicates 100.4 °C when a reference thermometer indicates 100.0 °C. Determine the indication error and correction.

Indication

I = 100.4 °C

Reference Value

R = 100.0 °C

Error

E = 100.4 − 100.0 = +0.4 °C

Correction

C = 100.0 − 100.4 = −0.4 °C

Corrected Indication

100.4 °C − 0.4 °C = 100.0 °C

Interpretation

The instrument reads 0.4 °C higher than the reference at this calibration point. The correction may be applied only within the justified calibration model and conditions, with uncertainty considered.

Calibration Intervals

Description

The time between successive calibrations should reflect the instrument's stability and risk of measurement error.

Selection Factors

  • Historical calibration results.

  • Instrument drift and stability.

  • Frequency and intensity of use.

  • Environmental and operating conditions.

  • Manufacturer recommendations.

  • Required measurement uncertainty.

  • Consequences of incorrect measurement.

  • Maintenance and adjustment history.

Review Methods

  • Historical trend analysis.

  • Intermediate checks using reference standards.

  • Control charts.

  • Risk-based interval adjustment.

Limitations And Precautions

  • Calibration does not guarantee that an instrument is suitable for every measurement task.

  • Calibration does not automatically include adjustment.

  • A calibration result is valid under the stated conditions and should not be assumed to apply to every operating environment.

  • A traceable calibration does not guarantee that measurement uncertainty is sufficiently small for the intended use.

  • Corrections should not be extrapolated beyond the calibrated range without justification.

  • A conformity statement requires defined requirements and an appropriate decision rule.

Reference

  • JCGM 200, International Vocabulary of Metrology (VIM).

  • ISO/IEC 17025:2017, General Requirements for the Competence of Testing and Calibration Laboratories.

  • ILAC P14, ILAC Policy for Measurement Uncertainty in Calibration.

  • EURAMET Calibration Guide No. 1, Determination of Measurement Uncertainties in Calibration.

Related Topic

  • SCI-REF-003-091

  • SCI-REF-003-092

  • SCI-REF-003-094

  • SCI-REF-003-097

Sources and verification

Sources are shown as provided in the project data. Not every source has yet been independently verified. Please report errors or missing citations.

  • Source verification pending; check an authoritative textbook, university, standards body, or peer-reviewed source.

Suggest a correction or source · More in Units & Measurement