Learning Objective
Define sensitivity, resolution and measurement range, distinguish these related specifications and interpret their effects on instrument selection and measurement quality.
Definition
Sensitivity is the change in instrument output divided by the corresponding change in input near a specified operating condition. Resolution is the smallest change in the measurand that can be distinguished or represented by the instrument. Measurement range is the interval of input values over which the instrument is specified to operate under stated conditions.
Scientific Background
Instrument specifications describe different aspects of performance. A highly sensitive instrument produces a relatively large output change for a small input change, but this does not guarantee that the change can be distinguished from noise. Resolution concerns distinguishable increments, while range describes the span of values that can be measured. These properties must be considered alongside accuracy, uncertainty, stability and response time.
Key Concepts
| Term | Explanation | Example |
|---|---|---|
Sensitivity | The slope of the input-output relationship, often expressed as Δoutput/Δinput over a stated region. | A temperature sensor whose output changes substantially for a small temperature change has higher sensitivity than one with a smaller output change, assuming comparable conditions. |
Resolution | The smallest input change that produces a distinguishable indication or output step under specified conditions. | A digital display increment of 0.01 °C indicates a display resolution of 0.01 °C, but not necessarily an accuracy of ±0.01 °C. |
Range | The minimum and maximum measurand values for which the instrument is specified to operate. | A pressure instrument specified from 0 to 500 kPa has a nominal measurement range of 0–500 kPa. |
Span | The difference between the upper and lower range limits. | For a 0–500 kPa range, the span is 500 kPa. |
Relationships And Equations
Sensitivity
S = Δy/Δx, where x is the input measurand and y is the instrument output, over the stated interval.
Span
Span = Upper range limit - Lower range limit.
Resolution Note
For a digital instrument, the least significant displayed increment may indicate display resolution; effective measurement resolution can be worse because of noise, stability and sensor limitations.
Worked Examples
| Problem | Calculation | Answer | Explanation |
|---|---|---|---|
An instrument output changes from 2.0 V to 3.5 V when its input changes from 10 units to 15 units. Find the average sensitivity over this interval. |
| 0.3 V per input unit | Average sensitivity is the output change divided by the input change over the stated interval. |
A thermometer is specified from -20 °C to 180 °C. Determine its span. |
| 200 °C | The span is the numerical difference between the upper and lower range limits. |
Performance Interpretation
High sensitivity can help detect small changes but may also make the system more responsive to noise or unwanted influences.
High display resolution does not guarantee low uncertainty or high accuracy.
An instrument should be operated within its specified range unless an approved method states otherwise.
Over-range conditions can cause invalid readings or damage, depending on instrument design.
Sensitivity may vary across the range, especially for nonlinear instruments.
Selection Criteria
Expected minimum and maximum values
Smallest change that must be detected
Required uncertainty and accuracy
Signal noise and environmental stability
Response time and measurement bandwidth
Need for overload protection
Common Errors
Treating sensitivity and resolution as synonyms
Assuming the last displayed digit is accurate
Confusing range with span
Using an instrument outside its specified range
Ignoring nonlinear sensitivity across the measurement range
Practical Applications
Selecting thermometers for narrow or broad temperature ranges
Choosing balances for different mass ranges and readability
Comparing pressure sensors and electronic measurement systems
Evaluating whether an instrument can detect a required process change
Knowledge Check
| Question | Answer | Explanation |
|---|---|---|
What does sensitivity describe? | How much the instrument output changes for a change in input. | It is commonly expressed as the slope of the calibration or response curve. |
Does a display resolution of 0.001 g mean the balance accuracy is ±0.001 g? | No. | Accuracy and uncertainty require additional performance information and calibration evidence. |