Learning Objective
Explain SEM scanning and signal detection and distinguish common SEM signals.
Major Systems
Electron source
Electromagnetic lenses and scanning coils
Specimen chamber and stage
Secondary-electron detector
Backscattered-electron detector
Vacuum system and control electronics
Working Principle
A focused electron beam is raster-scanned across the specimen. Interactions generate signals including secondary electrons and backscattered electrons. Signal intensity is mapped to beam position to form an image.
Signals
| Name | Information |
|---|---|
Secondary electrons | Often emphasize surface topography and fine surface detail. |
Backscattered electrons | Often provide atomic-number or compositional contrast, as well as topographic information depending on geometry. |
Characteristic X-rays | Can be analyzed using energy-dispersive X-ray spectroscopy to investigate elemental composition. |
Specimen Preparation
Specimens must tolerate the instrument's vacuum and electron beam. Nonconductive samples may require conductive coating or specialized low-vacuum/variable-pressure modes. Preparation depends on the material and analytical objective.
Applications
Fracture-surface examination
Microelectronics inspection
Particle morphology
Geological and biological materials
Elemental analysis when paired with suitable detectors
Limitations
Charging can distort images of insulating specimens.
Beam damage and contamination may occur.
A conventional SEM image is not automatically a quantitative 3D surface map.
Elemental analysis has detection limits and interaction-volume effects.
Safety
High-voltage and vacuum systems require trained operation and approved maintenance procedures.
Knowledge Check
Which signal is commonly used for surface topography?
Why may an insulating specimen charge under an electron beam?