Learning Objective
Explain quantum tunneling as the basis of STM and identify the conditions required for operation.
Components
Atomically sharp conductive tip
Conductive or semiconductive sample
Precision positioning scanner
Bias-voltage source
Tunneling-current amplifier
Feedback and vibration-control system
Working Principle
When a conductive tip is brought extremely close to a suitable surface and a bias voltage is applied, electrons can tunnel across the gap. The tunneling current depends strongly on tip–sample separation and also on electronic structure and bias conditions.
Imaging Modes
| Mode | Description |
|---|---|
Constant-current | Feedback adjusts tip height to maintain a selected tunneling current; the height variation is recorded. |
Constant-height | Tip height is held approximately fixed while current variation is recorded; this requires a sufficiently flat and stable surface. |
Requirements
A conductive or sufficiently semiconducting sample
A stable, clean tip and surface
Very low vibration and drift
Careful control of bias and current
Applications
Atomic-scale surface imaging
Surface electronic structure studies
Adsorbate and reconstruction analysis
Limitations
STM is not generally suitable for insulating samples. Apparent atomic contrast depends on electronic states as well as geometric topography.
Knowledge Check
Why does STM require a conductive path for tunneling current?
Does an STM image show only geometric height?