Fourier Transform Infrared (FTIR) Spectrometer

An FTIR spectrometer uses an interferometer to collect information from many infrared frequencies and applies a Fourier transform to produce an infrared spectrum.

Explanation

Learning Objective

Understand the interferometric operating principle of FTIR instruments and explain their data acquisition and sample analysis procedures.

Definition

A Fourier Transform Infrared spectrometer is an infrared analytical instrument that records an interferogram and mathematically transforms it into a spectrum showing signal intensity as a function of wavenumber.

Scientific Background

An interferometer combines optical paths to produce an intensity pattern that changes with optical path difference. This interferogram contains information about the infrared frequencies present in the signal.

Components

NameFunction

Broadband infrared source

Provides infrared radiation across the selected spectral range.

Beam splitter

Divides and recombines the infrared beam in the interferometer.

Fixed mirror

Reflects one portion of the beam along a fixed optical path.

Moving mirror

Changes the optical path difference during interferogram acquisition.

Sample compartment

Holds the sample in the selected measurement configuration.

Detector

Records the intensity variation associated with the interferogram.

Computer

Performs Fourier transformation and displays the resulting spectrum.

Working Principle

Principle

Interferometry followed by Fourier transformation

Explanation

The interferometer combines radiation traveling along different optical paths. As the optical path difference changes, the detector records an interferogram. A Fourier transform converts this signal into a frequency-domain spectrum.

Data Transformation

Interferogram → Fourier Transform → Infrared Spectrum

Important Concepts

  • Optical path difference

  • Interference

  • Interferogram

  • Fourier transformation

  • Phase correction

Measurement Procedure

  • Select a suitable sampling accessory, such as transmission or ATR.

  • Check the instrument condition and allow it to stabilize.

  • Collect a background interferogram without the sample.

  • Position the sample using the appropriate accessory.

  • Collect the sample interferogram.

  • Apply Fourier transformation and appropriate processing.

  • Inspect the spectrum for background interference and poor signal quality.

  • Compare spectral features with suitable reference data.

Advantages

  • Collects information from many infrared frequencies during an acquisition.

  • Can provide efficient data collection.

  • Offers good wavelength or wavenumber reproducibility when properly calibrated.

  • Can provide a favorable signal-to-noise ratio through signal averaging.

Limitations

  • Interferometer alignment and stability affect performance.

  • Background measurements may be required to compensate for atmospheric absorption.

  • Poor sample contact in ATR measurements can reduce spectral quality.

  • Data processing choices can influence the appearance of the spectrum.

Applications

  • Chemical identification

  • Polymer and material analysis

  • Pharmaceutical testing

  • Food quality analysis

  • Environmental sample characterization

  • Research into molecular interactions

Knowledge Check

QuestionAnswerExplanation

What mathematical operation converts an FTIR interferogram into a spectrum?

Fourier transformation.

The Fourier transform converts information recorded against optical path difference into frequency-domain information.

Example or application

A worked example has not yet been added.

Importance

FTIR instruments provide efficient infrared analysis and are widely used for chemical identification, material characterization and routine laboratory quality control.