Learning Objective
Understand the principle, components, sample introduction methods and analytical applications of atomic absorption spectroscopy.
Definition
An atomic absorption spectrometer is an analytical instrument that measures the absorption of characteristic radiation by free atoms, usually generated from a liquid sample.
Scientific Background
A sample is converted into free atoms in an atomization system. Atoms of a particular element absorb radiation at characteristic wavelengths corresponding to electronic transitions.
Components
| Name | Function |
|---|---|
Element-specific light source | Provides radiation characteristic of the element being measured, commonly using a hollow cathode lamp or an appropriate alternative. |
Atomizer | Converts sample constituents into free atoms. |
Nebulizer | Converts a liquid sample into an aerosol in flame systems. |
Monochromator | Selects the analytical wavelength and reduces unwanted radiation. |
Detector | Measures the intensity of transmitted radiation. |
Data system | Calculates absorbance and determines concentration using calibration data. |
Working Principle
Principle
Element-specific atomic absorption
Explanation
Radiation from an element-specific source passes through a region containing free atoms of the analyte. The atoms absorb characteristic radiation. The decrease in intensity is measured and related to the analyte concentration.
Measurement
Absorbance = log10(I0/I)
Quantification
Concentration is commonly determined using a calibration curve prepared from suitable standards.
Atomization Methods
| Name | Description |
|---|---|
Flame atomization | Uses a flame to produce free atoms from a nebulized sample. |
Electrothermal atomization | Uses an electrically heated graphite furnace to atomize a small sample quantity. |
Measurement Procedure
Select the element-specific analytical wavelength and lamp.
Prepare a blank and suitable calibration standards.
Prepare the sample using an appropriate digestion or dilution procedure.
Set up the selected atomization system.
Measure the blank and calibration standards.
Introduce the sample and record absorbance.
Determine concentration from the validated calibration relationship.
Apply dilution factors and quality-control checks.
Advantages
Good element-specific selectivity.
Useful for trace and major-element analysis, depending on configuration.
Well-established analytical methods are available for many elements.
Limitations
Many conventional AAS configurations measure elements sequentially.
Chemical and ionization interferences may affect results.
Sample digestion may be necessary for solid materials.
The method is limited to elements supported by the instrument and suitable analytical conditions.
Applications
Trace-metal analysis in water
Soil and sediment analysis
Food and agricultural testing
Pharmaceutical elemental analysis
Industrial material testing
Environmental monitoring
Maintenance And Safety
Follow approved flame and fuel-gas safety procedures.
Use appropriate ventilation and exhaust systems.
Handle acid digestion reagents using suitable laboratory precautions.
Keep burners and sample introduction components clean.
Verify calibration using appropriate quality-control samples.
Knowledge Check
| Question | Answer | Explanation |
|---|---|---|
Why is an element-specific lamp used in many AAS instruments? | It provides radiation at wavelengths characteristic of the element being analyzed. | Ground-state atoms of the target element absorb radiation at their characteristic electronic transition wavelengths. |