Learning Objective
Explain the combined plasma ionization and mass-analysis principle and identify the major applications and limitations of ICP-MS.
Definition
An inductively coupled plasma mass spectrometer is an analytical instrument that ionizes sample constituents in a plasma and measures the resulting ions using a mass analyzer.
Scientific Background
A liquid sample is commonly converted into an aerosol and introduced into an argon plasma. The plasma produces ions, which are transferred through an interface into a lower-pressure vacuum system. A mass analyzer separates ions according to mass-to-charge ratio.
Components
| Name | Function |
|---|---|
Sample introduction system | Delivers sample aerosol to the plasma. |
ICP torch | Generates the high-temperature argon plasma. |
Interface cones | Transfer ions from the atmospheric-pressure plasma region into the vacuum system. |
Ion optics | Guide and focus ions toward the mass analyzer. |
Collision or reaction cell | Reduces selected spectral interferences in instruments equipped with this system. |
Mass analyzer | Separates ions according to mass-to-charge ratio. |
Ion detector | Measures ion signals. |
Data system | Processes signals and calculates elemental or isotopic concentrations. |
Working Principle
Principle
Plasma ionization followed by mass-to-charge separation
Explanation
Sample elements are converted into ions in the plasma. The ions are transferred into the mass spectrometer, separated by mass-to-charge ratio and detected. Signal intensity is related to the amount of the corresponding ion after calibration and interference control.
Mass To Charge Ratio
m/z
Important Measurements
Elemental concentration
Isotope abundance
Isotope ratios in suitable configurations
Analysis Procedure
Prepare the sample using contamination-controlled procedures.
Prepare calibration standards, blanks and internal standards.
Check plasma gas, vacuum and sample introduction systems.
Select isotopes and operating conditions for the target elements.
Establish calibration and verify instrument response.
Analyze samples with appropriate quality-control checks.
Evaluate isobaric, polyatomic and matrix-related interferences.
Apply dilution and correction factors before reporting results.
Advantages
Very high sensitivity for many elements.
Broad multi-element capability.
Supports isotope measurements.
Can measure very low concentrations under validated conditions.
Limitations
Polyatomic and isobaric interferences can affect results.
Contamination can be significant at very low analyte levels.
High dissolved solids may cause matrix effects and instrument deposition.
Operating and maintenance costs can be substantial.
Applications
Trace-element analysis
Environmental monitoring
Geochemical analysis
Semiconductor material testing
Food and biological sample analysis
Isotope ratio research
Maintenance And Safety
Use clean sample handling procedures.
Follow plasma, gas and high-voltage safety requirements.
Maintain the interface cones and sample introduction components.
Monitor vacuum-system performance.
Use suitable blanks and certified reference materials.
Knowledge Check
| Question | Answer | Explanation |
|---|---|---|
What property is used by the mass analyzer to separate ions? | Mass-to-charge ratio. | Ions with different mass-to-charge ratios follow different trajectories or have different transmission conditions, depending on the analyzer. |