Definition
A colorimeter is an optical analytical instrument used to measure the intensity of color in a solution. It is commonly used to estimate the concentration of substances that absorb visible light or form colored products through chemical reactions.
Scientific Background
Colorimetric analysis is related to the Beer–Lambert law, which states that absorbance is proportional to concentration and optical path length under suitable conditions.
Working Principle
Light passes through a sample in a cuvette. A detector measures the transmitted light, and the instrument determines absorbance relative to an appropriate reference.
Formula
Name
Beer–Lambert Law
Expression
A = εlc
Variables
A
Absorbance
Ε
Molar absorptivity
L
Optical path length
C
Concentration
Components
Light source
Optical filter or wavelength selector
Sample holder
Cuvette
Photodetector
Signal processor
Display
Measurement Procedure
A suitable blank is used to establish the reference signal. The sample is placed in a compatible cuvette, and absorbance is measured at the selected wavelength. Concentration may then be estimated using a calibration curve.
Worked Examples
| Problem | Solution |
|---|---|
A calibration relationship is A = 0.20c, where c is expressed in mg/L. If the measured absorbance is 0.60, determine the concentration. | c = A / 0.20 = 0.60 / 0.20 = 3 mg/L. |
Types
| Name | Description |
|---|---|
Single-Beam Colorimeter | Measures the reference and sample signals separately. |
Digital Colorimeter | Uses electronic detection and digital display of measurement results. |
Applications
Water quality testing
Clinical chemistry
Food analysis
Educational laboratories
Chemical concentration measurement
Accuracy And Limitations
Accuracy depends on wavelength selection, cuvette cleanliness, sample clarity, calibration, and chemical stability. Turbid or strongly scattering samples may produce misleading absorbance readings.
Common Errors
Using an unsuitable blank
Fingerprints or contamination on cuvettes
Air bubbles in the sample
Incorrect wavelength selection
Using a calibration curve outside its valid range