Definition
A particulate matter monitor is an instrument that measures the mass concentration or estimated number concentration of airborne particles within defined size fractions.
Scientific Background
PM₁₀ refers to particles with aerodynamic diameters generally at or below 10 micrometres, while PM₂.₅ refers to particles generally at or below 2.5 micrometres. These categories are defined using aerodynamic size-selective sampling conventions.
Particle Categories
| Name | Description |
|---|---|
PM₁₀ | Inhalable particles with aerodynamic diameters up to approximately 10 µm. |
PM₂.₅ | Fine particles with aerodynamic diameters up to approximately 2.5 µm. |
Ultrafine particles | Particles commonly defined as having diameters below 0.1 µm; they are often characterized by number concentration. |
Instrument Types
| Name | Principle |
|---|---|
Gravimetric sampler | Collects particles on a filter and determines mass from the difference in filter weight. |
Beta attenuation monitor | Estimates particle mass from beta radiation attenuation. |
Tapered element oscillating microbalance | Estimates collected particle mass from changes in the oscillation frequency of a tapered element. |
Optical particle monitor | Estimates particle concentration from light scattering. |
Components
Size-selective inlet
Sampling pump
Particle collection filter or sensing chamber
Flow measurement system
Mass or optical detector
Data logger
Measurement Principle
The instrument separates particles according to aerodynamic size and determines their mass or optical response. Gravimetric methods determine collected mass directly, while continuous instruments use physical relationships to estimate concentration.
Formula
Name
Mass concentration
Expression
C = m / V
Variables
C
Particle mass concentration.
M
Collected particle mass.
V
Total volume of sampled air.
Worked Examples
| Problem | Solution |
|---|---|
A filter collects 120 µg of particles from 2 m³ of air. Calculate the average mass concentration. | C = 120 / 2 = 60 µg/m³. |
Measurement Units
µg/m³
mg/m³
particles/cm³ for number concentration
Accuracy And Limitations
Optical readings can be affected by particle composition, shape, and refractive properties.
High humidity can cause particles to absorb water and appear larger.
Filter methods require controlled weighing conditions.
Different monitoring technologies may not produce directly equivalent readings without appropriate correction or validation.
Calibration
Verify sampling flow rate.
Check particle size-selective inlet condition.
Use appropriate reference or comparison methods.
Follow filter conditioning and weighing procedures for gravimetric analysis.
Practical Applications
Urban air pollution monitoring
Industrial and construction-site assessment
Environmental research
Indoor and outdoor air studies
Long-term particulate pollution trends
Maintenance And Safety
Keep sampling inlets clean.
Replace filters according to the monitoring protocol.
Inspect pumps and flow sensors.
Handle collected filters using contamination-control procedures.
Knowledge Check
| Question | Answer |
|---|---|
What does PM₂.₅ represent? | Fine particulate matter with aerodynamic diameters up to approximately 2.5 µm. |
What is the SI-derived expression commonly used for particulate mass concentration? | Mass divided by sampled air volume, commonly reported in µg/m³. |