Definition
A turbidity meter measures the scattering or attenuation of light caused by particles suspended in water. Suspended solids measurement determines the mass of particulate material retained by a specified filtration method.
Scientific Background
Suspended particles include mineral sediment, organic matter, microorganisms, and other materials. They influence water clarity, light penetration, aquatic habitats, and water treatment processes.
Instrument Types
| Name | Principle |
|---|---|
Nephelometric turbidity meter | Measures light scattered by particles, commonly at an angle of 90 degrees to the incident beam. |
Attenuation turbidity meter | Measures reduction in transmitted light caused by suspended particles. |
Optical backscatter sensor | Measures light scattered back toward a detector. |
Gravimetric suspended solids system | Uses filtration, drying, and weighing to determine particulate mass. |
Components
Light source
Optical sample chamber
Photodetector
Signal processor
Sample cell
Calibration standards
Measurement Concepts
| Name | Definition |
|---|---|
Turbidity | An optical property related to the scattering and attenuation of light by suspended material. |
Total suspended solids (TSS) | The mass concentration of material retained by a specified filter under a defined analytical method. |
Formula
Name
Total suspended solids
Expression
TSS (mg/L) = [(m2 - m1) × 1000] / V
Variables
M1
Mass of the prepared filter before filtration, in mg.
M2
Mass of the filter and retained dried residue, in mg.
V
Filtered sample volume, in litres.
Worked Examples
| Problem | Solution |
|---|---|
A filter gains 12 mg after filtering 0.5 L of water. Calculate TSS. | TSS = (12 × 1000) / 0.5 = 24,000 mg/L. |
Measurement Units
Turbidity: NTU or FNU, depending on optical method
TSS: mg/L
Accuracy And Limitations
Turbidity is not a direct measurement of suspended solids mass.
Particle size, shape, and color influence optical response.
Air bubbles can cause falsely elevated turbidity readings.
Filter type and analytical procedure influence TSS results.
Calibration
Use appropriate turbidity reference standards.
Clean sample cells and optical surfaces.
Follow the specified filtration and drying procedure for TSS.
Use duplicate samples and suitable quality-control checks.
Practical Applications
Drinking-water source monitoring
River and lake assessment
Wastewater treatment
Sediment transport studies
Environmental pollution monitoring
Maintenance And Safety
Prevent fingerprints and scratches on optical cells.
Mix samples consistently where appropriate.
Use suitable precautions for potentially contaminated water.
Maintain filter, balance, and instrument quality-control records.
Knowledge Check
| Question | Answer |
|---|---|
Does turbidity directly equal TSS? | No. Turbidity is an optical measurement, while TSS is a mass concentration. |
Which optical arrangement is common in nephelometric turbidity measurement? | A detector positioned approximately 90 degrees to the incident light beam. |