Definition
Zero adjustment concerns the instrument response at or near the lower reference point of its measurement range. Span refers to the difference between the upper and lower range values, and span calibration evaluates the instrument response across that range.
Key Terms
Zero
The lower reference point or designated zero indication of a measurement system.
Span
The difference between the upper and lower range values.
Zero Error
The difference between the instrument indication and the reference value at the zero or lower reference point.
Span Error
An error affecting the response across the measurement range, often observed by comparing the upper-range response with its reference value.
Formulas
Span
Span = Upper Range Value - Lower Range Value
Measurement Error
Error = Instrument Indication - Reference Value
Conceptual Example
Range
0 to 100 units
Span
100 units
Reference Points
| Reference Value | Purpose |
|---|---|
0 units | Evaluate the lower reference response. |
100 units | Evaluate the upper reference response. |
Comparison
| Feature | Focus |
|---|---|
Zero evaluation | Instrument response at the lower reference point. |
Span evaluation | Instrument response across the measurement range. |
Important Note
Zero and span adjustment are changes to instrument settings. Calibration itself is the comparison and evaluation process; adjustment should only be performed when justified and permitted by the applicable procedure.
Practical Applications
Pressure transmitters.
Temperature measurement systems.
Electronic sensors.
Industrial process indicators.
Knowledge Check
| Question | Answer |
|---|---|
What does span represent? | The difference between the upper and lower measurement range values. |